Stereotactic vacuum-assisted breast biopsy: Comparison between 11- and 8-gauge needles

Published:October 06, 2017DOI:https://doi.org/10.1016/j.ejso.2017.07.011

      Abstract

      Purpose

      The 11-gauge (11G) stereotactic vacuum-assisted breast biopsy (VABB) showed a better profile than 14G-VABB in terms of feasibility, safety, microcalcification sampling, and accuracy. Underestimation rates were significantly lower with 11G-VABB than with 14G-VABB. Thus, the introduction of an even larger needle at the VABB procedure could reduce this rate further. The purpose of this study was to compare the overall performance of stereotactic VABB with 8G and 11G needles.

      Materials and methods

      Four hundred and three VABBs performed between July 2012 and February 2015 at the Breast Diagnostic Unit of Careggi Hospital in Florence were retrospectively analyzed; 197 were performed with 11G-VABB and 206 with 8G-VABB. Lesions were classified according to mammographical patterns in microcalcifications, architectural distortions, or opacities, and all biopsy targets were classified according to BIRADS classification as BIRADS III, IV or V. Data were collected on radiological classification of targets, imaging presentation, procedure time, number of specimens per procedure, and microcalcification retrieval on histological findings. Surgery was always performed when high-risk or malignant lesions (B3 or B5) were detected; the final diagnosis was made on surgical pathology.

      Results

      Compared to VABB with an 11G needle, 8G-VABB allows a reduction in the time needed to complete the procedure (20.6 versus 27.4, P < 0.00001) and the number of specimens collected per lesion (21.6 versus 12.2, P < 0.00001). Moreover, 8G-VABB resulted in the same diagnostic accuracy, and the underestimation rates were comparable between the two groups for both B3 and DCIS lesions.

      Conclusions

      The 8G needle should be considered as a valid alternative option in VABB for breast lesions.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to European Journal of Surgical Oncology
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Burbank F.
        • Parker S.H.
        • Fogarty T.J.
        Stereotactic breast biopsy: improved tissue harvesting with the Mammotome.
        Am Surg. 1996; 62: 738-744
        • Burbank F.
        Stereotactic breast biopsy of atypical hyperplasia and ductal carcinoma in situ lesions: improved accuracy with directional, vacuum-assisted biopsy.
        Radiology. 1997; 202: 843-847
        • Kettritz U.
        • Rotter K.
        • Schreer I.
        • Murauer M.
        • Schulz-Wendtland R.
        • Peter D.
        • et al.
        Stereotactic vacuum-assisted breast biopsy in 2874 patients: a multicenter study.
        Cancer. 2004; 100: 245-251
        • Meyer J.E.
        • Smith D.N.
        • DiPiro P.J.
        • Denison C.M.
        • Frenna T.H.
        • Harvey S.C.
        • et al.
        Stereotactic breast biopsy of clustered microcalcifications with a directional, vacuum-assisted device.
        Radiology. 1997; 204: 575-576
        • Philpotts L.E.
        • Shaheen N.A.
        • Carter D.
        • Lange R.C.
        • Lee C.H.
        Comparison of rebiopsy rates after stereotactic core needle biopsy of the breast with 11-gauge vacuum suction probe versus 14-gauge needle and automatic gun.
        AJR Am J Roentgenol. 1999; 172: 683-687
        • Jackman R.L.
        • Burbank F.
        • Parker S.H.
        • Evans 3rd, W.P.
        • Lechner M.C.
        • Richardson T.R.
        • et al.
        Atypical ductal hyperplasia diagnosed et stereotactic breast biopsy: improved reliability with 14-gauge, directional, vacuum-assisted biopsy.
        Radiology. 1997; 204: 485-488
        • Darling M.L.
        • Smith D.N.
        • Lester S.C.
        • Kaelin C.
        • Selland D.L.
        • Denison C.M.
        • et al.
        Atypical ductal hyperplasia and ductal carcinoma in situ as revealed by large-core needle breast biopsy: results of surgical excision.
        AJR Am J Roentgenol. 2000; 175: 1341-1346
        • Liberman L.
        • Smolkin J.H.
        • Dershaw D.D.
        • Morris E.A.
        • Abramson A.F.
        • Rosen P.P.
        Calcification retrieval at stereotactic, 11-gauge, directional. vacuum-assisted breast biopsy.
        Radiology. 1998; 208: 251-260
        • Hahn M.
        • Okamgba S.
        • Scheler P.
        • Freidel K.
        • Hoffmann G.
        • Kraemer B.
        • et al.
        Vacuum-assisted breast biopsy: a comparison of 11-gauge and 8-gauge needles in benign breast disease.
        World J Surg Oncol. 2008; 6: 51
        • Diebold T.
        • Hahn T.
        • Solbach C.
        • Rody A.
        • Balzer J.O.
        • Hansmann M.L.
        • et al.
        Evaluation of the stereotactic 8G vacuum-assisted breast biopsy in the histologic evaluation of suspicious mammography findings (BI-RADS IV).
        Invest Radiol. 2005; 40: 465-471
        • Lee K.E.
        • Kim H.H.
        • Shin H.J.
        • Cha J.H.
        Stereotactic biopsy of the breast using a decubitus table: comparison of histologic underestimation rates between 11- and 8-gauge vacuum-assisted breast biopsy.
        Springerplus. 2013; 2: 551
        • Lifrange E.
        • Dondelinger R.F.
        • Quatresooz P.
        • Vandevorst G.
        • Colin C.
        Stereotactict breast biopsy with an 8-gauge, directional, vacuum assisted probe:initial experience.
        Eur Radiol. 2002; 12: 2180-2187
        • BI-RADS®–Mammography
        Fourth edition American College of Radiology (ACR) breast imaging reporting and data system Atlas (BI-RADS® Atlas).
        American College of Radiology, Reston, VA2003
        • Venkataraman S.
        • Dialani V.
        • Gilmore H.L.
        • Mehta T.S.
        Stereotactic core biopsy: comparison of 11 gauge with 8 gauge vacuum assisted breast biopsy.
        Eur J Radiol. 2012; 81: 2613-2619
        • Brem R.F.
        • Schoonjans J.M.
        • Goodman S.N.
        • Nolten A.
        • Askin F.B.
        • Gatewood O.M.
        Nonpalpable breast cancer: percutaneous diagnosis with 11- and 8- gauge stereotactic vacuum-assisted biopsy devices.
        Radiology. 2001; 219: 793-796
        • Lomoschitz F.M.
        • Helbich T.H.
        • Rudas M.
        • Pfarl G.
        • Linnau K.F.
        • Stadler A.
        • et al.
        Stereotactic 11-gauge vacuum-assisted breast biopsy: influence of number of specimens on diagnostic accuracy.
        Radiology. 2004; 232: 897-903
        • Pfarl G.
        • Helbich T.H.
        • Riedl C.C.
        • Wagner T.
        • Gnant M.
        • Rudas M.
        • et al.
        Stereotactic 11-gauge vacuum-assisted breast biopsy: a validation study.
        AJR. 2002; 179: 1503-1507